Magnitude, Phase – Atec Agilent-N5230A User Manual

Page 16

background image

16

Table 8. Dynamic Accuracy (specification)


Accuracy of the test port input power reading relative to the reference input power level.
Dynamic accuracy is verified with the following measurements:

• Compression over frequency

• IF linearity at a single frequency of 1.195 GHz using a reference level of –20 dBm

for an input power range of 0 to –110 dBm

Dynamic Accuracy 0.045 GHz

Magnitude

0.01

0.1

1

10

-120

-110

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Test port power (dBm)

N5230A Option 140/145/146/240/245/246

Accuracy (dB)

-10 dBm at 0.045 GHz

-20 dBm at 0.045 GHz

-30 dBm at 0.045 GHz

-40 dBm at 0.045 GHz

Phase

0.1

1

10

100

-120

-110

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Test port power (dBm)

N5230A Option 140/145/146/240/245/246

Accuracy (degrees)

-10 dBm at 0.045 GHz

-20 dBm at 0.045 GHz

-30 dBm at 0.045 GHz

-40 dBm at 0.045 GHz