11

Agilent 8719ES or 8720ES with Option 007 or 085

Calibration kit: 85052D 3.5-mm with fixed loads
Cables: 85131F 3.5-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

Corrected measurement port specifications

Frequency range (GHz)

0.05 to 0.5

0.5 to 2

2 to 8

8 to 20

Directivity

42 dB

42 dB

38 dB

36 dB

Source match

37 dB

37 dB

30 dB

28 dB

Load match

42 dB

41 dB

36 dB

34 dB

Reflection tracking

±(0.006 dB + .02 dB/°C)

±(0.01 dB + .03 dB/°C)

±(0.03 dB + .03 dB/°C)

±(0.31dB + .04 dB/°C)

Transmission tracking

±(0.018 dB + .02 dB/°C)

±(0.019 dB + .03 dB/°C)

±(0.080 dB + .03 dB/°C)

±0.141 dB + .04 dB/°C)

Magnitude

Phase

Magnitude

Phase

Measurement uncertainty

Reflection measurements

Transmission measurements