beautypg.com

System performance, continued – Atec Agilent-8722ES User Manual

Page 12

background image

16

Agilent 8719ES, 8720ES, or 8722ES Option 012

Option 012 adds RF loops that provide direct access
to the A and B samplers in the port 1 and port 2
receivers. This allows transmission measurements
that bypass the receiver coupler for improved signal-
to-noise and sensitivity. The system is capable of
antenna measurements to –110 dBm at 40 GHz,
and filter rejection measurements to 120 dB. Use
of multiple antennae provides improved signal-to-
noise for free-space transmission and reflection
measurements. The RF loops can also be used to
integrate components into the test set. Adding a
20 dB attenuator increases the test port 0.1 dB com-
pression level to +30 dBm. With front panel jumpers
installed, the system operates as a standard system
and meets standard instrument specifications.

Supplemental characteristics (Option 012)

Frequency range (GHz)

0.05 to 0.5 0.5 to 2

2 to 8

8 to 20

20 to 40

Compression

1

Test Port

1,2

13 dB

10 dB

10 dB

10 dB

10 dB

Compression

1

Direct Sampler Input

–5 dBm

–5 dBm

–5 dBm

–5 dBm

–5 dBm

Average
Noise Floor

2

–120 dBm –120 dBm –120 dBm –118 dBm –113 dBm

Receiver
Dynamic Range

115 dB

115 dB

115 dB

113 dB

108 dB

System performance, continued

Option 012 test set block diagram

Option 400 and 012 test set block diagram

Measure filter rejection to –120 dB

Transfer
Switch

Source

Port 1

Port 2

Samplers

A

R

B

R channel
jumper

Switch
Splitter

Source

Port 1

Port 2

Samplers

A

R

R2

B

R channel
jumper

Samplers

A

B

Measure amplifier output to +43 dBm

Samplers

A

B

16 dB more sensitivity for antenna test. Improved signal-
to-noise for free-space materials test.

Samplers

A

B

1. Input power level that causes 0.1 dB compression in the receiver.
2. 10 Hz IF BW.