beautypg.com

Measurement throughput summary, Full frequency band sweep time (ms) – Atec Agilent-8722ES User Manual

Page 16

background image

20

Full frequency band sweep time (ms)

1

Number of Points

Measurement

51

201

401

1601

(Swept mode/Stepped mode)

Single band sweep

(10 to 12 GHz)

Uncorrected

27 / 134

65 / 492

116 / 970

419 / 3836

One-port calibration

2

27 / 134

65 / 492

116 / 970

419 / 3836

Two-port calibration

3

80 / 492

158 / 1034

259 / 2010

866 / 7885

8719E full sweep

(0.05 to 13.5 GHz)

Uncorrected

484 / 597

553 / 1014

614 / 1490

926 / 4336

One-port calibration

2

484 / 597

553 / 1014

614 / 1490

926 / 4336

Two-port calibration

3

996 / 1222

1133 / 2069

1259 / 3057

1876 / 8892

8720E full sweep

(0.05 to 20 GHz)

Uncorrected

449 / 581

538 / 1017

598 / 1490

900 / 4335

One-port calibration

2

449 / 581

538 / 1017

598 / 1490

900 / 4335

Two-port calibration

3

930 / 1192

1106 / 2172

1227 3053

1826 / 8892

8722E full sweep

(0.05 to 40 GHz)

Uncorrected

570 / 731

651 / 1162

707 / 1690

961 / 4519

One-port calibration

2

570 / 731

651 / 1162

707 / 1690

961 / 4519

Two-port calibration

3

1168 / 1162

1333 / 2367

1444 / 3439

1949 / 9269

Time-domain conversion

4

13

42

87

380

15

41

81

350

GPIB data transfer

5

Binary (Internal)

35

15

20

54

IEEE754 floating point format
32 bit

11

18

26

78

64 bit

13

24

40

134

ASCII

33

105

203

781

Measurement throughput summary

1. All values are typical.
2. S11 one-port calibration, with a 6 kHz IF bandwidth. Includes system retrace time. time-domain gating is assumed off.
3. S21 measurement with full two-port calibration, using a 6 kHz IF bandwidth. Includes system retrace time and RF switching time.

Time-domain gating is assumed off.

4. Option 010 only, gating and error-correction are off. Does not include sweep time.
5. Measured with an HP Omnibook 7100 266 PentiumII computer.