17

Agilent 8719ES, 8720ES, and 8722ES Option 089

Option 089 adds frequency offset mode, allowing
the receiver to be offset from the source frequency.
This allows direct conversion loss measurement of
mixers without need for a reference mixer. RF and
IF frequencies must be within the specified operat-
ing range of the instrument. This test set modifica-
tion adds an internal reference channel switch and
deletes the reference channel output. Firmware
guides the user through the test setup. When not
in frequency offset mode, the system operates as
a standard system and meets standard instrument
specifications.

Supplemental characteristics (Option 089)

8719ES/8720ES

8722ES

Reference (R) Input Level
Minimum

–34 dBm

–34 dBm

Maximum (for 0.1 dB compression) –7 dBm

–12 dBm

Maximum (damage level)

0 dBm

0 dBm

Note: To utilize full instrument receiver dynamic range,
measure test signal at port 2. This configuration requires
a power splitter and reference mixer to provide a phase
lock signal to the R channel input.

Analyzer’s guided test setup display

Option 089 test set block diagram

Mechanical
transfer switch

R channel
input

LO source

Source

Samplers

Reference

switch

Mixer

under test

A

R

RF

IF

LO

B

Filter