Atec Agilent-8722ES User Manual
Page 13
17
Agilent 8719ES, 8720ES, and 8722ES Option 089
Option 089 adds frequency offset mode, allowing
the receiver to be offset from the source frequency.
This allows direct conversion loss measurement of
mixers without need for a reference mixer. RF and
IF frequencies must be within the specified operat-
ing range of the instrument. This test set modifica-
tion adds an internal reference channel switch and
deletes the reference channel output. Firmware
guides the user through the test setup. When not
in frequency offset mode, the system operates as
a standard system and meets standard instrument
specifications.
Supplemental characteristics (Option 089)
8719ES/8720ES
8722ES
Reference (R) Input Level
Minimum
–34 dBm
–34 dBm
Maximum (for 0.1 dB compression) –7 dBm
–12 dBm
Maximum (damage level)
0 dBm
0 dBm
Note: To utilize full instrument receiver dynamic range,
measure test signal at port 2. This configuration requires
a power splitter and reference mixer to provide a phase
lock signal to the R channel input.
Analyzer’s guided test setup display
Option 089 test set block diagram
Mechanical
transfer switch
R channel
input
LO source
Source
Samplers
Reference
switch
Mixer
under test
A
R
RF
IF
LO
B
Filter