System performance, continued – Atec Agilent-8722ES User Manual
Page 10
Agilent 8719ET and 8720ET
with 3.5 mm test ports
Standard or Option 004
Calibration kit: 85052B 3.5-mm with sliding loads
Cables: 85131F 3.5-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)
System dynamic range
0.05 to 0.84 GHz
102 dB
0.84 to 20 GHz
104 dB
Power range:
Standard:
–10 to +10 dBm
Option 004: –65 to +10 dBm
14
System performance, continued
Magnitude
Phase
Measurement uncertainty
Reflection measurements
Magnitude
Phase
Magnitude
Phase
1. One-port or enhanced-response calibration
2. Enhanced-response or response-only calibration
3. Option 004 may degrade uncorrected source match as much as 2 dB.
Transmission measurements
Corrected measurement port specifications (3.5 mm)
Frequency range (GHz)
0.050 to 0.5 0.5 to 2
2 to 8
8 to 20
Reflection Measurements
1
Directivity
48 dB
48 dB
44 dB
44 dB
Source match
40 dB
40 dB
33 dB
31 dB
Load match
One-port Cal
22 dB
22 dB
22 dB
17 dB
Enhanced Reflection Cal 22 dB
22 dB
22 dB
17 dB
Reflection
tracking
±(0.006
dB ±(0.006
dB ±(0.006
dB ±(0.008
dB
+0.02 dB/°C) +0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C)
Transmission measurements
2
Source match
Enhanced Response Cal 40 dB
40 dB
33 dB
31 dB
Response Only Cal
16 dB
20 dB
14 dB
11 dB
Transmission tracking
Enhanced Response Cal ±(0.014 dB
±(0.012 dB
±(0.027 dB
±(0.050 dB
+0.02 dB/°C) +0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C)
Response Only Cal
±(0.109 dB
±(0.069 dB
±(0.137 dB
±(0.339 dB
+0.02 dB/°C) +0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C)
Load match
22 dB
22 dB
22 dB
17 dB