background image

Agilent 8719ET and 8720ET
with 3.5 mm test ports

Standard or Option 004

Calibration kit: 85052B 3.5-mm with sliding loads
Cables: 85131F 3.5-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

System dynamic range

0.05 to 0.84 GHz

102 dB

0.84 to 20 GHz

104 dB

Power range:
Standard:

–10 to +10 dBm

Option 004: –65 to +10 dBm

14

System performance, continued

Magnitude

Phase

Measurement uncertainty

Reflection measurements

Magnitude

Phase

Magnitude

Phase

1. One-port or enhanced-response calibration
2. Enhanced-response or response-only calibration
3. Option 004 may degrade uncorrected source match as much as 2 dB.

Transmission measurements

Corrected measurement port specifications (3.5 mm)

Frequency range (GHz)

0.050 to 0.5 0.5 to 2

2 to 8

8 to 20

Reflection Measurements

1

Directivity

48 dB

48 dB

44 dB

44 dB

Source match

40 dB

40 dB

33 dB

31 dB

Load match

One-port Cal

22 dB

22 dB

22 dB

17 dB

Enhanced Reflection Cal 22 dB

22 dB

22 dB

17 dB

Reflection

tracking

±(0.006

dB ±(0.006

dB ±(0.006

dB ±(0.008

dB

+0.02 dB/°C) +0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C)

Transmission measurements

2

Source match

Enhanced Response Cal 40 dB

40 dB

33 dB

31 dB

Response Only Cal

16 dB

20 dB

14 dB

11 dB

Transmission tracking

Enhanced Response Cal ±(0.014 dB

±(0.012 dB

±(0.027 dB

±(0.050 dB

+0.02 dB/°C) +0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C)

Response Only Cal

±(0.109 dB

±(0.069 dB

±(0.137 dB

±(0.339 dB

+0.02 dB/°C) +0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C)

Load match

22 dB

22 dB

22 dB

17 dB