beautypg.com

Measurement uncertainty, Corrected measurement port specifications (2.4 mm) – Atec Agilent-8722ES User Manual

Page 11

background image

Agilent 8722ET with 2.4 mm test ports

Standard or Option 004

Calibration kit: 85056A 2.4-mm with sliding loads
Cables: 85133F 2.4-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

System dynamic range

0.05 to 0.84 GHz

98 dB

0.84 to 8 GHz

102 dB

8 to 20 GHz

100 dB

20 to 40 GHz

89 dB

Power range
0.05 to 20 GHz: –15 to 0 dBm
20 to 40 GHz: –15 to –5 dBm
Option 004
0.05 to 20 GHz: –70 to 0 dBm
20 to 40 GHz: –70 to –5 dBm

15

Magnitude

Phase

Measurement uncertainty

Reflection measurements

Transmission measurements

Magnitude

Phase

Magnitude

Phase

1. One-port or enhanced-response calibration
2. Enhanced-response or response-only calibration
3. Option 004 may degrade uncorrected source match as much as 2 dB.

Corrected measurement port specifications (2.4 mm)

Frequency range (GHz)

0.050 to 2

2 to 8

8 to 20

20 to 40

Reflection Measurements

1

Directivity

42 dB

42 dB

42 dB

38 dB

Source match

41 dB

38 dB

38 dB

33 dB

Load match

One-port Cal

20 dB

20 dB

20 dB

15 dB

Enhanced Reflection Cal 20 dB

20 dB

20 dB

15 dB

Reflection tracking

±(0.005 dB

±(0.010 dB ±(0.010 dB

±(0.021 dB

+0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C) +0.06 dB/°C)

Transmission measurements

2

Source match

Enhanced Response Cal 41 dB

38 dB

38 dB

33 dB

Response Only Cal

16 dB

14 dB

11 dB

10 dB

Transmission tracking

Enhanced Response Cal ±(0.018 dB

±(0.026 dB

±(0.031 dB

±(0.069 dB

+0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C) +0.06 dB/°C)

Response Only Cal

±(0.109 dB

±(0.172 dB

±(0.241 dB

±(0.475 dB

+0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C) +0.06 dB/°C)

Load match

20 dB

20 dB

20 dB

15 dB