Measurement uncertainty, Corrected measurement port specifications (2.4 mm) – Atec Agilent-8722ES User Manual
Page 11
Agilent 8722ET with 2.4 mm test ports
Standard or Option 004
Calibration kit: 85056A 2.4-mm with sliding loads
Cables: 85133F 2.4-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)
System dynamic range
0.05 to 0.84 GHz
98 dB
0.84 to 8 GHz
102 dB
8 to 20 GHz
100 dB
20 to 40 GHz
89 dB
Power range
0.05 to 20 GHz: –15 to 0 dBm
20 to 40 GHz: –15 to –5 dBm
Option 004
0.05 to 20 GHz: –70 to 0 dBm
20 to 40 GHz: –70 to –5 dBm
15
Magnitude
Phase
Measurement uncertainty
Reflection measurements
Transmission measurements
Magnitude
Phase
Magnitude
Phase
1. One-port or enhanced-response calibration
2. Enhanced-response or response-only calibration
3. Option 004 may degrade uncorrected source match as much as 2 dB.
Corrected measurement port specifications (2.4 mm)
Frequency range (GHz)
0.050 to 2
2 to 8
8 to 20
20 to 40
Reflection Measurements
1
Directivity
42 dB
42 dB
42 dB
38 dB
Source match
41 dB
38 dB
38 dB
33 dB
Load match
One-port Cal
20 dB
20 dB
20 dB
15 dB
Enhanced Reflection Cal 20 dB
20 dB
20 dB
15 dB
Reflection tracking
±(0.005 dB
±(0.010 dB ±(0.010 dB
±(0.021 dB
+0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C) +0.06 dB/°C)
Transmission measurements
2
Source match
Enhanced Response Cal 41 dB
38 dB
38 dB
33 dB
Response Only Cal
16 dB
14 dB
11 dB
10 dB
Transmission tracking
Enhanced Response Cal ±(0.018 dB
±(0.026 dB
±(0.031 dB
±(0.069 dB
+0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C) +0.06 dB/°C)
Response Only Cal
±(0.109 dB
±(0.172 dB
±(0.241 dB
±(0.475 dB
+0.03 dB/°C) +0.03 dB/°C) +0.04 dB/°C) +0.06 dB/°C)
Load match
20 dB
20 dB
20 dB
15 dB