beautypg.com

Atec Agilent-8722ES User Manual

Page 3

background image

7

Agilent 8722ES with 2.4 mm test ports

Standard, Options 400, 012

1

, 089, or any combina-

tion of these options.

Calibration kit: 85056D, 2.4-mm with broadband
loads
Cables: 85133F 2.4-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

System dynamic range

0.05 to 0.84 GHz

67 dB

0.84 to 8 GHz

93 dB

8 to 20 GHz

91 dB

20 to 40 GHz

80 dB

2

Corrected measurement port specifications

Frequency range (GHz)

0.05 to 2

2 to 8

8 to 20

20 to 40

Directivity

42 dB

42 dB

34 dB

26 dB

Source match

40 dB

40 dB

30 dB

23 dB

Load match

42 dB

42 dB

34 dB

26 dB

Reflection tracking

±(0.006 dB + 0.03 dB/°C)

±(0.029 dB + 0.03 dB/°C)

±(0.029 dB + 0.04 dB/°C)

±(0.080 dB + 0.06 dB/°C)

Transmission tracking

±(0.022 dB + 0.03 dB/°C)

±(0.034 dB + 0.03 dB/°C)

±(0.116 dB + 0.04 dB/°C)

±(0.372 dB + 0.06 dB/°C)

Maximum output power
0.05 to 20 GHz: –5 dBm
20 to 40 GHz: –10 dBm

Magnitude

Phase

Magnitude

Phase

Measurement uncertainty

Reflection measurements

Transmission measurements

1. Internal test set not bypassed.
2. 3 dB less with Option 012.