beautypg.com

System performance, continued – Atec Agilent-8722ES User Manual

Page 2

background image

6

Agilent 8719ES and 8720ES
with 3.5 mm test ports

Standard, Options 400, 012

1

, 089, or any combina-

tion of these options.

Calibration kit: 85052D, 3.5-mm with broadband
loads
Cables: 85131F 3.5-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

System dynamic range

0.05 to 0.84 GHz

77 dB

0.84 to 20 GHz

100 dB

System performance, continued

Magnitude

Phase

Magnitude

Phase

Measurement uncertainty

Reflection measurements

Transmission measurements

Corrected measurement port specifications

Frequency range (GHz)

0.05 to 0.5

0.5 to 2

2 to 8

8 to 20

Directivity

42 dB

42 dB

38 dB

36 dB

Source match

37 dB

37 dB

31 dB

28 dB

Load match

42 dB

42 dB

38 dB

36 dB

Reflection tracking

±(0.006 dB + .02 dB/°C)

±(0.006 dB + .03 dB/°C)

±(0.006 dB + .03 dB/°C)

±(0.009 dB + .04 dB/°C)

Transmission tracking

±(0.028 dB + .02 dB/°C)

±(0.03 dB + .03 dB/°C)

±(0.096 dB + .03 dB/°C)

±(0.158 dB + .04 dB/°C)

Maximum output power
+5 dBm

1. Internal test set not bypassed.