beautypg.com

Specifications and characteristics – Atec Agilent-8722ES User Manual

Page 14

background image

18

Agilent 8719ES, 8720ES, and 8722ES

Description

Code

Frequency
Range

8719ES

0.05 to 13.51 GHz

S

8720ES

0.05 to 20.05 GHz

S

8722ES

0.05 to 40.05 GHz

S

Accuracy (at 23° C ±3° C)

±10 ppm

S

Stability

0° to 55° C

±7.5 ppm

C

Option 1D5

±0.05 ppm

C

Per year (aging)

±3 ppm

C

Option 1D5

±0.5 ppm

C

Resolution

1 Hz

S

System
Maximum input level

10 dBm

S

Damage level (test port)

30 dBm or 40V dc

C

Trace noise

1

Magnitude (RMS of peak-to-peak value)

0.05 to 13.5 GHz

<0.03 dB

C

13.5 to 20 GHz

<0.04 dB

C

20 to 40 GHz

<0.15 dB

C

Phase (RMS of peak-to-peak value)

0.05 to 13.5 GHz

<0.3°

C

13.5 to 20 GHz

<0.4°

C

20 to 40 GHz

<1.5°

C

Spectral purity
Harmonics at maximum output level

<–15 dBc

C

Phase noise

to 60 kHz from carrier at 2 GHz

<–55 dBc

T

to 60 kHz from carrier at 20 GHz

<–35 dBc

T

Nonharmonic spurious signals

at 100 kHz offset

<–40 dBc

T

at 200 kHz offset

<–45 dBc

T

at >200 kHz offset

<–65 dBc

T

Description

Code

Power
Power range

8719ES/8720ES

75 dB

C

8722ES

0.05 to 20 GHz

70 dB

C

20 to 40 GHz

65 dB

C

Maximum output power

8719ES/8720ES (standard, Options 085, 400)

+5 dBm

C

8719ES/8720ES (Option 007)

+10 dBm

C

8722ES (standard, Options 085, 400)

0.05 to 20 GHz

–5 dBm

C

20 to 40 GHz

–10 dBm

C

8722ES (Option 007)

0.05 to 20 GHz

0 dBm

C

20 to 40 GHz

–5 dBm

C

Resolution

0.01 dB

S

Level accuracy (at 5 dB below maximum output power)

8719ES/8720ES

±2 dB

S

8722ES

±3 dB

S

Power sweep range

8719ES/8720ES

20 dB

S

8719ES/8720ES

30 dB

T

8722ES

0.05 to 20 GHz

15 dB

S

8722ES

0.05 to 20 GHz

25 dB

T

8722ES

20 to 40 GHz

10 dB

S

8722ES

20 to 40 GHz

20 dB

T

Power linearity

–5 dB from reference:

8719ES/8720ES

±0.35

S

8722ES

0.05 to 20 GHz

±0.35

S

8722ES

20 to 40 GHz

±0.60

S

+5 dB from reference:

8719ES/8720ES

±0.35

S

8722ES

0.05 to 20 GHz

±0.35

S

8722ES

20 to 40 GHz

N/A

–10 dB from reference:

8719ES/8720ES/8722ES

±0.60

S

+10 dB from reference:

8719ES/8720ES

±1.0

S

8722ES

N/A

Test reference power levels:

8719ES/8720ES (standard, Options 085, 400)

–5 dBm

S

8719ES/8720ES (Option 007)

0 dBm

S

8722 ES (standard, Options 085, 400)

–10 dBm

S

8722ES (Option 007)

–5 dBm

S

Specifications and characteristics

Frequency range (GHz)

Description

0.05 to 0.84

0.84 to 2

2 to 8

8 to 20

20 to 40

Code

System
Dynamic range

2

8719ES/8720ES (standard, Options 085, 400)

77 dB

100 dB

100 dB

100 dB

S

8719ES/8720ES (Option 007)

82 dB

105 dB

105 dB

105 dB

S

8722ES (standard

3

, Options 085, 400)

67 dB

93 dB

93 dB

91 dB

80 dB

4

S

8722ES (Option 007)

72 dB

98 dB

98 dB

96 dB

85 dB

S

Receiver 0.1 dB compression input level

5

13 dBm

10 dBm

10 dBm

10 dBm

10 dBm

T

Codes
(S) Specification: warranted performance. Specifications include guardbands

to account for expected statistical performance distribution, measurement
uncer-tainties, and changes in performance due to environmental conditions.

(C) Characteristic: a performance parameter that the product is expected to meet

before it leaves the factory, but that is not verified in the field and is not covered
by the warranty. A characteristic includes the same guardbands as a specification.

(T) Typical (typ.): expected performance of an average unit which does not include

guardbands. It is not covered by the product warranty.

(N) Nominal (nom.): a general, descriptive term that does not imply a level of

performance. It is not covered by the product warranty.

1. Trace noise is defined as variation of a high signal level trace due to noise. The

value given represents a noise variation that is the RMS of the peak-to-peak value
as measured in a 3 kHz IF bandwidth in the fast sweep mode (STEP SWP OFF).

2. The dynamic range specifications apply to transmission measurements using

10 Hz IF BW and response and isolation correction or full two-port correction.
Dynamic range is limited by the maximum test port power and the receiver’s
noise floor.

3. With 85133E flexible cable on test port.
4. 77 dB for Option 085.
5. For Option 012 direct sampler input compression levels, refer to page 16.