Atec Agilent-8722ES User Manual
Page 15
19
Agilent 8719ET, 8720ET, and 8722ET
Description
Code
Frequency
Range
8719ET
0.05 to 13.51 GHz
S
8720ET
0.05 to 20.05 GHz
S
8722ET
0.05 to 40.05 GHz
S
Accuracy (at 23° C ±3° C)
±10 ppm
S
Stability
0° to 55° C
±7.5 ppm
C
Option 1D5
±0.05 ppm
C
Per year (aging)
±3 ppm
C
Option 1D5
±0.5 ppm
C
Resolution
1 Hz
S
System
Maximum input level
+5 dBm transmission S
+10 dBm reflection
S
Damage level (test port)
27 dBm or 0V dc
C
Trace noise
2
Magnitude (RMS of peak-to-peak value)
0.05 to 13.5 GHz
<0.03 dB
C
13.5 to 20 GHz
<0.04 dB
C
20 to 40 GHz
<0.15 dB
C
Phase (RMS of peak-to-peak value)
0.05 to 13.5 GHz
<0.3°
C
13.5 to 20 GHz
<0.4°
C
20 to 40 GHz
<1.5°
C
Spectral purity
Harmonics at maximum output level
<–15 dBc
C
Phase noise
to 60 kHz from carrier at 2 GHz
<–55 dBc
T
to 60 kHz from carrier at 20 GHz
<–35 dBc
T
Nonharmonic spurious signals
at 100 kHz offset
<–40 dBc
T
at 200 kHz offset
<–45 dBc
T
at >200 kHz offset
<–65 dBc
T
Description
Code
Power
Power range
8719ET/8720ET
20 dB
C
8722ET
0.05 to 20 GHz
15 dB
C
20 to 40 GHz
10 dB
C
Power range (Option 004)
8719ET/8720ET
75 dB
C
8722ET
0.05 to 20 GHz
70 dB
C
20 to 40 GHz
65 dB
C
Maximum output power
8719ET/8720ET
+10 dBm
C
8722ET
0.05 to 20 GHz
0 dBm
C
20 to 40 GHz
–5 dBm
C
Resolution
0.01 dB
S
Level accuracy (at 5 dB below maximum output power)
8719ET/8720ET
±2 dB
S
8722ET
±3 dB
S
Power sweep range
8719ET/8720ET
20 dB
S
8719ET/8720ET
30 dB
T
8722ET
0.05 to 20 GHz
15 dB
S
8722ET
0.05 to 20 GHz
25 dB
T
8722ET
20 to 40 GHz
10 dB
S
8722ET
20 to 40 GHz
20 dB
T
Power linearity
–5 dB from reference:
8719ET/8720ET
±0.35
S
8722ET
0.05 to 20 GHz
±0.35
S
8722ET
20 to 40 GHz
±0.60
S
+5 dB from reference:
8719ET/8720ET
±0.35
S
8722ET
0.05 to 20 GHz
±0.35
S
8722ET
20 to 40 GHz
N/A
–10 dB from reference:
8719ET/8720ET/8722ET
±0.60
S
+10 dB from reference:
8719ET/8720ET
±1.0
S
8722ET
N/A
Test reference power levels:
8719ET/8720ET (standard, Option 004)
0 dBm
S
8722 ET (standard, Option 004)
–5 dBm
S
Frequency range (GHz)
Description
0.05 to 0.84
0.84 to 2
2 to 8
8 to 20
20 to 40
Code
System
Dynamic range
3
8719ET/8720ET (standard, Options 004)
102 dB
104 dB
104 dB
104 dB
S
8722D (standard
4
, Option 400)
98 dB
102 dB
102 dB
100 dB
89 dB
S
Receiver 0.1 dB compression input level
13 dBm
10 dBm
10 dBm
10 dBm
10 dBm
T
Codes
(S) Specification: warranted performance. Specifications include guardbands
to account for expected statistical performance distribution, measurement
uncertainties, and changes in performance due to environmental conditions.
(C) Characteristic: a performance parameter that the product is expected to meet
before it leaves the factory, but that is not verified in the field and is not covered
by the warranty. A characteristic includes the same guardbands as a specification.
(T) Typical (typ.): expected performance of an average unit which does not include
guardbands. It is not covered by the product warranty.
(N)Nominal (nom.): a general, descriptive term that does not imply a level of
performance. It is not covered by the product warranty.
1. Does not apply to 8722ET.
2. Trace noise is defined as variation of a high signal level trace due to noise. The
value given represents a noise variation that is the RMS of the peak-to-peak value
as measured in a 3 kHz IF bandwidth in the fast sweep mode (STEP SWP OFF).
3. The dynamic range specifications apply to transmission measurements using
10 Hz IF BW and response and isolation correction or full two-port correction.
Dynamic range is limited by the maximum test port power and the receiver’s
noise floor.
4. With 85133E flexible cable on test port.