System performance, continued – Atec Agilent-8722ES User Manual
Page 4
8
Agilent 8719ES Option 400, 8720ES Option 400,
and 8722ES Option 400 with 3.5 mm test ports
using TRL
Includes instruments with Options 012
1
and/or 089.
Calibration kit: 85052C, 3.5-mm for TRL
Cables: 85131F 3.5-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)
System dynamic range
Frequency range
8719ES/8720ES
8722ES
0.05 to 0.084 GHz
77 dB
67 dB
0.084 to 8 GHz
100 dB
93 dB
8 to 20 GHz
100 dB
91 dB
20 to 40 GHz
—
80 dB
2
Maximum output power
8719ES/8720ES: +5 dBm
8722ES (0.05 to 20 GHz): –5 dBm
(20 to 40 GHz): –10 dBm
System performance, continued
Corrected measurement port specifications
8719ES/8720ES
Frequency range (GHz)
Option 400
0.05 to 0.5
0.5 to 2
2 to 8
8 to 20
Directivity
48 dB
48 dB
50 dB
50 dB
Source match
40 dB
40 dB
50 dB
50 dB
Load match
48 dB
48 dB
50 dB
50 dB
Reflection tracking
±(0.006 dB + 0.02 dB/°C)
±(0.006 dB + 0.03 dB/°C)
±(0.005 dB + 0.03 dB/°C)
±(0.005 dB + 0.04 dB/°C)
Transmission tracking
±(0.013 dB + 0.02 dB/°C)
±(0.017 dB + 0.03 dB/°C)
±(0.06 dB + 0.03 dB/°C)
±(0.019 dB + 0.04 dB/°C)
8722ES Option 400
Frequency range (GHz)
0.05 to 2
2 to 8
8 to 20
20 to 26.5
Directivity
48 dB
50 dB
50 dB
50 dB
Source match
40 dB
50 dB
50 dB
50 dB
Load match
48 dB
50 dB
50 dB
50 dB
Reflection tracking
±(0.006 dB + 0.03 dB/°C)
±(0.005 dB + 0.03 dB/°C)
±(0.005 dB + 0.04 dB/°C)
±(0.005 dB + 0.06 dB/°C)
Transmission tracking
±(0.017 dB + 0.03 dB/°C)
±(0.013 dB + 0.03 dB/°C)
±(0.016 dB + 0.04 dB/°C)
±(0.023 dB + 0.06 dB/°C)
Magnitude
Phase
Magnitude
Phase
Measurement uncertainty
Reflection measurements
Transmission measurements
1. Internal test set not bypassed.
2. 3 dB less with Option 012.