YSI MultiLab 4010P-1 User Manual
Page 71

MultiLab 4010-1
Settings
ba76141e03
03/2014
71
ORP
The following settings for ORP measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the key.
Dissolved oxygen
The following settings for D.O. measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the key.
Conductivity
The following settings for conductivity measurements are reset to the default
settings with the Reset function:
Setting
Default settings
mV resolution
0.1
Man. temperature
25 °C
Setting
Default settings
Calibration interval
180 d
Measured parameter
D.O. concentration (mg/l)
Salinity (value)
0.0
Salinity (function)
off
Resolution
0.1
DO % Saturation local
off
Setting
Default settings
Calibration interval
150 d
Measured parameter
ϰ
Cell constant (c)
Depending on the connected mea-
suring cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set)
0.100 cm
-1
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of the
linear temperature compensation
2.000 %/K
Multiplier for TDS
1.00