7.2.6 Calibration data
Displaying the
calibration record
The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration/Calibration data storage / Display. To open the Calibration menu,
press the key in the measured value display.
9.
Wait for the AutoRead measurement to be completed (audio signal).
When calibrating in water vapor-saturated air (Cal):
The calibration record is displayed and output to the interface.
When calibrating via comparison measurement (Comp):
Set the concentration value with <> <> and accept it with
The calibration record is displayed and output to the interface.
10.
Switch to the measured value display with .
The calibration record is automatically transmitted to the interface
after calibrating.
O
2
DO Adjust
7.92
mg/l
Comp
03.04.2013
08:00
760 mmHg
25.1 °C