YSI MultiLab 4010P-1 User Manual

Page 32

background image

pH value

MultiLab 4010-1

32

ba76141e03

03/2014

Continuing with

three- to five-point

calibration

13. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with

.
The calibration display for the next buffer appears (voltage display).

14. If necessary, finish the calibration procedure as a two-point calibration

with .
The calibration record is displayed.

pH

0.0

24.8

°C

AutoCal YSI

7.000

01.07.2011

08:00

AR

Buffer

mV

15. Thoroughly rinse the sensor with deionized water.

16. Immerse the sensor in the next buffer solution.

17. Start the measurement with

.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

18. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with

.
The calibration display for the next buffer appears (voltage display).

19. If necessary, use to finish the calibration.

The calibration record is displayed.
or
Switch to calibration with the next buffer with

.

pH

180.0

24.8

°C

AutoCal YSI

4.00

01.07.2011

08:00

AR

Buffer

mV