YSI MultiLab 4010P-1 User Manual
Page 58

Conductivity
MultiLab 4010-1
58
ba76141e03
03/2014
Calibration
evaluation
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Calibration record
(USB output)
Calibration /
Calibration data
storage / Output to
RS232/USB
-
Outputs the calibration data storage
to the interface
Menu item
Setting/
function
Description
Display
Calibration record
Cell constant [cm
-1
]
+++
Within the range
0.450 ... 0.500 cm
-1
Error
Error
Outside the range
0.450 ... 0.500 cm
-1
Error elimination (see section 14
W
CALIBRATION Cond
03.04.2013 07:43:33
4310
Ser. no. 09250033
Cell constant 0.476 1/cm 25.0 °C
Sensor
+++
This manual is related to the following products: