YSI MultiLab 4010P-1 User Manual
Page 64

Settings
MultiLab 4010-1
64
ba76141e03
03/2014
Calibration /
Calibration data
storage / Number
of cal. points
1
2
1-point calibration
or
2-point calibration
Calibration /Cali-
bration interval
1 ... 180 ...
999 d
Calibration interval for the IDS D.O.
sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
Calibration /Com-
parison meas.
on
off
Enables to adjust the measured value
with the aid of a comparison measure-
ment, e.g. Winkler titration.
For details, see section 7.2 C
Calibration /
Sensor cap coef-
ficients
K1
...
K5
KC
Enter the cap coefficients here after
exchanging the sensor cap.
For details, see section 9.3.2 E
Salinity/Sal cor-
rection
on
off
Manual salt content correction for con-
centration measurements.
Salinity/Salinity
0.0 ... 70.0
Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if the
manual salt content correction is
switched on.
Resolution
0.1
1
Set a high or low resolution.
The setting of the resolution is sepa-
rately stored for each measured
parameter.
DO % Saturation
local
on
off
DO % Saturation local is a procedure
that takes the local air pressure into
account for each saturation measure-
ment.
For details, see section 9.3.3 DO %
S
Reset
-
Resets all sensor settings to the deliv-
ery condition (see section 9.6.1
R
Menu item
Possible
setting
Description