9.5.2 Data storage
This menu contains all functions to display, edit and erase stored measured
values.
9.5.3 Automatic Stability control
The automatic Stability control (AutoRead) function continuously checks the
stability of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
You can activate or switch off the automatic Stability control function (see
section 9.5 S
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
9.5.4 Automatic shutoff function
The instrument has an automatic shutoff function in order to save the batteries
(see section 9.5.1 S
, page 68). The automatic shutoff function switches
off the meter if no key is pressed for an adjustable period.
The automatic shutoff function is not active
if the power pack is connected
if a USB-B cable is connected
if the Automatic data storage function is active, or with automatic data trans-
mission
9.5.5 Display illumination
The meter automatically switches off the display illumination if no key is
pressed for 20 seconds.
The illumination is switched on with the next keystroke again.
System /Reset
-
Resets the system settings to the
default values.
For details, see section 9.6.2 R
Menu item
Setting
Description
Detailed information on the memory functions of the
MultiLab 4010-1 are given in section 10 D