5 calibration points, Calibration points – YSI MultiLab 4010P-1 User Manual
Page 36

pH value
MultiLab 4010-1
36
ba76141e03
03/2014
5.2.5 Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
22. Set the nominal buffer value for the measured temperature with
<><>.
23. Accept the calibration value with .
The calibration display for the next buffer appears (voltage display).
24. If necessary, use
The calibration record is displayed.
or
Continue calibrating using the next buffer with .
After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
Calibration
Determined values
Displayed calibration data
1-point
Asy
Zero point = Asy
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asy
Slp.
Zero point = Asy
Slope = Slp.
3-point to 5-
point
Asy
Slp.
Zero point = Asy
Slope = Slp.
The calibration line is calculated
by linear regression.
You can display the slope in the units, mV/pH or % (see section
9.1.1 S