Carrying out a QSC

control calibration

A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably
changed (after some normal calibrations).

You can carry out QSC control calibrations at greater intervals than normal cali-
brations.

1 QSC scale

The double arrow on the QSC scale indicates the current sensor
evaluation

1

25.0

7.007

pH

°C

0

14

01.02.2014

08:00

Info

USB output

1.

Open the menu for measurement settings with

.

2.

In the QSC menu, select Control calibration with <><>.
The calibration display appears. AutoCal QSC-Kit is displayed as the
buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other
buffers, you will have no valid QSC control calibration.

3.

Follow the user guide.
The calibration is carried out like a normal three-point calibration. As
soon as the three-point calibration has been successfully carried out
you can decide whether to accept or discard the calibration as the QSC
control calibration.