6 reset, 1 resetting the measurement settings, Reset – YSI MultiLab 4010P-1 User Manual
Page 70: Resetting the measurement settings, On (see sec- tion 9.6.1 r, Esetting, Measurement, Settings, Section 9.6.1 r, Tings

Settings
MultiLab 4010-1
70
ba76141e03
03/2014
You can also generally switch on the display illumination (see section 9.5.1
S
9.6
Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
9.6.1 Resetting the measurement settings
pH
The following settings for pH measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the key.
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer YSI
Calibration interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Unit for slope
0.001
mV resolution
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
Single-point calibration
off