5 2-point calibration, Point calibration – YSI MultiLab 4010P-1 User Manual
Page 49

MultiLab 4010-1
Dissolved oxygen
ba76141e03
03/2014
49
7.2.5 2-point calibration
Calibration in water vapor-saturated air (Cal) and calibration via Comparison
meas. according to the Winkler method (Comp) are described jointly here.
6.
Wait for the AutoRead measurement to be completed (audio signal).
When calibrating in water vapor-saturated air (Cal):
The calibration record is displayed and output to the interface.
When calibrating via comparison measurement (Comp):
Set the concentration value with <> <> and accept it with
The calibration record is displayed and output to the interface.
7.
Switch to the measured value display with .
1.
Connect the IDS D.O. sensor to the meter.
2.
Check the selected calibration type and selected number of calibration
points in the sensor setting menu and, if necessary, change them.
3.
Place the IDS D.O. sensor in a solution that does not contain any dis-
solved oxygen.
A solution not containing any dissolved oxygen can be prepared by
dissolving approx. 8 to 10 g sodium sulfite (Na
2
SO
3
) in 500 ml tap-
water. Carefully mix the solution. It may take up to 60 minutes until
the solution is free of oxygen.
4.
Start the calibration with
Cal (calibration in water vapor-saturated air) or Comp (calibration via
Comparison meas. according to the Winkler method) is displayed.
The calibration point 1 (0%) is displayed.
O
2
DO Adjust
7.92
mg/l
Comp
03.04.2013
08:00
760 mmHg
25.1 °C