5.2.5 Calibration points

Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:

22. Set the nominal buffer value for the measured temperature with

<><>.

23. Accept the calibration value with

.

The calibration display for the next buffer appears (voltage display).

24. If necessary, use to finish the calibration.

The calibration record is displayed.
or
Continue calibrating using the next buffer with

.

After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.

The calibration line is determined by linear regression.

Calibration

Determined values

Displayed calibration data

1-point

Asy

Zero point = Asy

Slope = Nernst slope
(-59.2 mV/pH at 25 °C)

2-point

Asy
Slp.

Zero point = Asy

Slope = Slp.

3-point to 5-
point

Asy
Slp.

Zero point = Asy

Slope = Slp.

The calibration line is calculated
by linear regression.

You can display the slope in the units, mV/pH or % (see section
9.1.1 S

ETTINGS

FOR

P

H

MEASUREMENTS

, page 59).