beautypg.com

3 sensor status sxx, Senso – YSI DIQ/S 182 XT-4-PR User Manual

Page 77

background image

System 182 XT-4

Operation

4 - 11

ba76027e01

02/2012

4.4.3

Sensor status Sxx

In the SENSOR STATUS Sxx menu you can prompt the following
information on a sensor:

 Type and series number

 Software status

 Error and info messages ("log book")

 Data of the last calibration (only for sensors that can be calibrated)

Fig. 4-8

SENSOR STATUS Sxx

Log book

The log book is a list of messages related to a sensor. If there is a new
message the info or error symbol in the header of the measured value
display flashes:

If there are info and error messages present at the same time, the error
symbol

h

flashes (more important).

The flashing of the info or error symbol only stops after all detailed
message texts in the log book have been opened and marked with a
tick (). The operating sequence to do this is described on page 4-13.

Note
The error symbol also flashes if the operational voltage or
communication in the system are faulty. You can check this in the
PROPERTIES / SYSTEM STATUS menu (see section 4.4.1).

h

Flashes if a new or unacknowledged error message is present
in the log book that requires immediate action.

i

Flashes if a new or unacknowledged informational message is
present in the log book.