8 using diq modules (accessories), 1 diq/jb, 2 diq/chv – YSI DIQ/S 182 XT-4-PR User Manual

Page 45: Using diq modules (accessories) -31 3.8.1, Diq/jb -31, Diq/chv -31, Section 3.8

background image

System 182 XT-4

Installation

3 - 31

ba76027e01

02/2012

3.8

Using DIQ modules (accessories)

Note
The various application possibilities of the DIQ modules are shown by
means of examples in section 3.13.

3.8.1

DIQ/JB

The DIQ/JB module is a passive branching module and can be used for
the following purposes:

 To extend the IQ S

ENSOR

N

ET

SNCIQ(/UG) cable and the SACIQ

sensor connection cable, e.g. to connect a sensor that is located
farther away to the DIQ/S 182 XT-4-PR.

 Connection of another sensor to the DIQ/S 182 XT-4-PR if all

SENSORNET connections are already in use. Example of the
connection of four sensors to the DIQ/S 182 XT-4-PR.

Fig. 3-22 DIQ/JB open.

The DIQ/JB module has seven potential free terminals. To extend or
branch lines, connect the three IQ S

ENSOR

N

ET

wires to each other

one-to-one at any terminal.

 green <-> green

 red <-> red

 black/filler stranded wire <-> black/filler stranded wire.

3.8.2

DIQ/CHV

The DIQ/CHV module is a valve module for the automatic relay-
controlled compressed air-driven cleaning function in the System
182 XT-4. It provides four additional potential free terminals to branch
(extend) interface lines. For each sensor that is to have compressed air
cleaning a DIQ/CHV is required.