8 using diq modules (accessories), 1 diq/jb, 2 diq/chv – YSI DIQ/S 182 XT-4-PR User Manual
Page 45: Using diq modules (accessories) -31 3.8.1, Diq/jb -31, Diq/chv -31, Section 3.8
System 182 XT-4
Installation
3 - 31
ba76027e01
02/2012
3.8
Using DIQ modules (accessories)
Note
The various application possibilities of the DIQ modules are shown by
means of examples in section 3.13.
3.8.1
DIQ/JB
The DIQ/JB module is a passive branching module and can be used for
the following purposes:
To extend the IQ S
ENSOR
N
ET
SNCIQ(/UG) cable and the SACIQ
sensor connection cable, e.g. to connect a sensor that is located
farther away to the DIQ/S 182 XT-4-PR.
Connection of another sensor to the DIQ/S 182 XT-4-PR if all
SENSORNET connections are already in use. Example of the
connection of four sensors to the DIQ/S 182 XT-4-PR.
Fig. 3-22 DIQ/JB open.
The DIQ/JB module has seven potential free terminals. To extend or
branch lines, connect the three IQ S
ENSOR
N
ET
wires to each other
one-to-one at any terminal.
green <-> green
red <-> red
black/filler stranded wire <-> black/filler stranded wire.
3.8.2
DIQ/CHV
The DIQ/CHV module is a valve module for the automatic relay-
controlled compressed air-driven cleaning function in the System
182 XT-4. It provides four additional potential free terminals to branch
(extend) interface lines. For each sensor that is to have compressed air
cleaning a DIQ/CHV is required.