3 sensor monitoring, Senso, See section 6.4.3 – YSI DIQ/S 182 XT-4-PR User Manual
Page 106
Relay outputs
System 182 XT-4
6 - 16
ba76027e01
02/2012
6.4.3
Sensor monitoring
Function
The Sensor monitoring function enables the monitoring of sensor errors
and the maintenance condition.
In order to set up the Sensor monitoring function for a relay output, the
relay output must be linked with a sensor (see section 6.2.1).
Settings
Note
Preferably set up the relay output as an opener for monitoring
functions.
Sensor messages include errors and information that are registered by
the sensor.
Special
sensor errors
The special sensor errors are sensor-dependent. Details of this are
given in the component operating manual of the respective sensor.
General
sensor errors
Setting
Selection
Explanation
Errors
All
All sensor errors (special ones
and general ones) are
monitored and can prompt a
relay action.
Special
Special sensor errors are
monitored and can prompt a
relay action.
Off
Sensor errors are not
monitored.
Mainten. condition
On
Off
Switching on and off the
maintenance condition is
monitored and can prompt a
relay action.
Action
Open
For all functions of the Sensor
monitoring, the relay action is
set to Open.
Init
This can prompt a relay action for a short time,
depending on the starting behavior of the system
----
Invalid measured value, or defective sensor
Error
Communication with sensor interrupted
OFL
Measuring range undercut or exceeded (overflow)