6 profibus error diagnosis, 7 profibus error elimination, Profibus error diagnosis -12 – YSI DIQ/S 182 XT-4-PR User Manual
Page 128: Profibus error elimination -12
PROFIBUS connection
System 182 XT-4
7 - 12
ba76027e01
02/2012
7.6
PROFIBUS error diagnosis
A simple diagnosis of the PROFIBUS communication is possible via
the measured value and status display (status online/offline, see
section 4.2).
7.7
PROFIBUS error elimination
Note
Here you will find causes and actions to take of errors concerning the
PROFIBUS communication only. General errors of the IQ S
ENSOR
N
ET
system are dealt with in chapter 9 W
Data transmission
between the Profibus
master and
DIQ/S 182 XT-4-PR is
faulty
The PLC does not
receive any plausible
input data
Cause
Remedy
– Incorrect wiring of the
PROFIBUS cable on the
connector
– Check/change the connections
(see section 7.2)
– Use checklist according to
– Incorrect setting of the
address
– Check/change the setting of
the address (see section 7.3)
– Incorrect protocol
– Check the version of the GSD
file
– Adapt the protocol
– DIQ/S 182 XT-4-PR
defective
– Send DIQ/S 182 XT-4-PR to
YSI
Cause
Remedy
– Input data and output data
are not consistent
– When programming the PLC,
define input data and output
data as consistent over the
entire data length
– The data interpretation of the
PLC is not correct
– Observe the data alignment of
the PLC data representation. If
necessary, exchange the high-
order and low-order bytes word
by word