4 properties menu, 1 overview, Properties – YSI DIQ/S 182 XT-4-PR User Manual
Page 74: Overview -8
Operation
System 182 XT-4
4 - 8
ba76027e01
02/2012
4.4
PROPERTIES menu
4.4.1
Overview
Pressing
g
switches from the measured value display to the
PROPERTIES menu.
In the PROPERTIES menu, you can prompt calibration data and
various information on system components. Moreover, you can switch
the maintenance condition of a sensor on and off.
Fig. 4-7
PROPERTIES menu
Submenus/functions
Maintenance S
XX
This function switches the maintenance condition
of a sensor on and off (see section 4.4.2)
Sensor status Sxx
Here you can prompt the following data of a sensor (see
section 4.4.3):
– Type, series number and software version
– Data of the last calibration
– Error and info messages
System status
Here you can prompt the following data of the system:
– Series number and software version of the DIQ/S 182 XT-4-PR
– Status of the operational voltage (
= OK /
= faulty)
– Status of communication (
= OK /
= faulty)
– Current states of interfaces
If the operational voltage or communication are faulty, the flashing
error symbol
h
appears in the header of the measured value
display. The error symbol
h
automatically disappears when the
malfunction has been eliminated.
List of components
Here is a list of all connected sensors.