System 182 XT-4
Relay outputs
6 - 11
ba76027e01
02/2012
Characteristic curves
Through the selection of the Start value and End value, the proportional
output can be operated with a positive or negative characteristic curve.
Positive characteristic curve:
Select the
End value to be greater than the Start value.
The turn-on duration or frequency increases with an increasing
measured value (see page 6-12).
Negative characteristic curve:
Select the
End value to be smaller than the Start value.
The turn-on duration or frequency decreases with an increasing
measured value (see page 6-13).
The maximum values for the pulse width or switching frequency are
assigned to the End value value and the minimum values for turn-on
duration or frequency are assigned to the Start value value.
Fig. 6-6
Positive characteristic curve
Fig. 6-7
Negative characteristic curve
Measured value
Switching frequency f or
Pulse
v
width
End value
Start value
Max.
Min.
Measured value
Start value
End value
Max.
Min.
Switching frequency f or
Pulse
v
width
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (128 pages)
- 5200A (121 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (143 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IFL 700 IQ (46 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)