Dissolved oxygen calibration, Do calibration in % saturation, L. see section 6.2.2 dissolved oxygen calibration – YSI 5200 User Manual
Page 65: 2 dissolved oxygen calibration
YSI Incorporated
63
5200 Recirculating System Monitor
6.2.2 Dissolved
Oxygen
Calibration
This procedure calibrates dissolved oxygen. Calibrating any one option (% or mg/L) automatically
calibrates the other.
Salininty 0-45ppt
mg/L
%Sat
Display Actual Value % (CAL Value via Math)
Scrolling Text*1
Display Actual Value mg/L
Scrolling Text*2
Calibration
Complete
Yes
Calibration
If Cond turned
Off Display
Question
No
Alt 100 X 100 Ft
Calibration
Complete
Get UserValue
Figure 6-1 DO Calibration Menu
DO Calibration in % Saturation
1. Go to the DO calibrate screen as described in Section 6.2.1 Accessing the Calibrate Screen.
2. If your probe module contains a conductivity probe and it is Enabled, skip to Step 3. If your probe
module does not contain a conductivity probe (or the conductivity probe is not Enabled), enter
the salinity of the sample stream.
NOTE: The salinity compensation range for DO is 0–45 ppt.