Calibration/storage cup (5562 only), Calibration/storage cup installation, Instrument/cable connection – YSI 5200 User Manual
Page 33: Alibration, Torage, Only, Nstrument, Able, Onnection, 8 instrument/cable connection
YSI Incorporated
31
5200 Recirculating System Monitor
CAUTION: Do not touch the membrane surface.
8. Screw the membrane cap onto the sensor moderately tight. A small amount of electrolyte should
overflow.
9. IMPORTANT: To prevent the formation of a bulge in the membrane (which can slow sensor
response), partially unscrew the membrane cap then retighten.
10. Screw the probe sensor guard on moderately tight.
4.7 Calibration/Storage Cup (5562 only)
The YSI 5562 Probe module is supplied with a convenient calibration/storage cup. This cup is an ideal
container for calibration of the different sensors, minimizing the amount of solution needed. Refer to
Section 6 Calibration.
4.7.1
Calibration/Storage Cup Installation
1. Remove probe sensor guard, if already installed.
2. Ensure that an o-ring is installed in the o-ring groove on the threaded end of the probe module
body.
3. Screw the calibration/storage cup on the threaded end of the probe module and securely tighten.
NOTE: Do not over tighten as this could cause damage to the threaded portions.
Figure 4-9 Calibration/Storage Cup Installation
4.8 Instrument/Cable
Connection
Attach the probe cable to the instrument as follows:
1. Line up the pins and guides on the cable with the holes and indentations on the cable connector at
the bottom of the YSI 5200 instrument.
O-ring