Temperature system, 2 temperature system, Ysi incorporated – YSI 5200 User Manual
Page 39: Press the enter, Key to enter the temp system menu, Temperature system menu selections are shown below
YSI Incorporated
37
5200 Recirculating System Monitor
Parameter
Set Point
High Set
Point #1
High Set
Point #2
Low Set
Point #1
Low Set
Point #2
High Alarm
Low Alarm
ORP
+400 mV
+425 mV
---
+375 mV
---
+435 mV
+300 mV
Temperature 25
o
C or 77
o
F 26
o
C or
78.8
o
F
--- 24
o
C or
75.2
o
F
--- 26.5
o
C or
79.7
o
F
23.5
o
C or
74.3
o
F
pH 7.00
7.50
---
6.50
---
8.00
6.00
Specific Conductance
(range dependent;
0-200 us, 0-2000 us,
0-20 ms, 0-200 ms)
(0-200 us)
160.0 us
(0-2000 us)
1600 us
(0-20 ms)
16.00 ms
(0-200 ms)
160.0 ms
(0-200 us)
180.0 us
(0-2000 us)
1800 us
(0-20 ms)
18.00 ms
(0-200 ms)
130.0 ms
--- (0-200
us)
140.0 us
(0-2000 us)
1400 us
(0-20 ms)
14.00 ms
(0-200 ms)
140.0 ms
--- (0-200
us)
190.0 us
(0-2000 us)
1900 us
(0-20 ms)
19.00 ms
(0-200 ms)
190.0 ms
(0-200 us)
130.0 us
(0-2000 us)
1300 us
(0-20 ms)
13.00 ms
(0-200 ms)
130.0 ms
Dissolved Oxygen
8.3 mg/L
8.5 mg/L
8.7 mg/L
8.1 mg/L
7.9 mg/L
8.9 mg/L
7.7 mg/L
5.3.2 Temperature
System
Temperature is explained in step-by-step detail as an example. Once you are familiar with the temperature
set up procedure, other parameters can be set up in a similar manner.
From the Parameters menu, press the [e] or [f] key until the Temp System option appears.
T e m p S y s t e m
E n t e r t o S e l e c t
Press the Enter [
] key to enter the Temp System menu.
T e m p S y s t e m
E N A B L E D ? Y e s
If the Temperature System is disabled (as indicated by No appearing on the display), press the [e] or [f]
key to select Yes.
T e m p S y s t e m
E N A B L E D ? Y e s
Press the Enter [
] key to confirm that the option is Enabled and enter the Temp System menu.
S e t P o i n t M e n u
E n t e r t o S e l e c t
Temperature System menu selections are shown below.