Calibration procedures, Accessing the calibrate screen, Alibration – YSI 5200 User Manual
Page 64: Rocedures, 2 calibration procedures
YSI Incorporated
62
5200 Recirculating System Monitor
6.2 Calibration
Procedures
6.2.1
Accessing the Calibrate Screen
1. From the Run screen, press the Enter [
] key, then the [e] or [f] key until the Parameters
option appears.
P a r a m e t e r s
E n t e r t o S e l e c t
2. Press the Enter [
] key to select Parameters.
D O S y s t e m
E n t e r t o S e l e c t
3. Press the [e] or [f] key until the parameter you want to calibrate appears. Conductivity is used
as an example below.
C o n d S y s t e m
E n t e r t o S e l e c t
4. Press the Enter [
] key to display the Enabled screen.
C o n d S y s t e m
E N A B L E D ? N o
5. If the parameter is disabled (as indicated by No appearing on the display), press the [e] or [f]
key to select Yes.
C o n d S y s t e m
E N A B L E D ? Y e s
6. Press the Enter [
] key.
C a l i b r a t i o n
E n t e r t o S e l e c t
7. Press the Enter [
] key to select Calibration.
8. Follow the instructions in the following sections for the specific parameter that you are
calibrating.