Laser Troubleshooting

Problem

The laser will not start.

Probable Cause

On water-cooled lasers, bubbles may have developed in the cooling system. This will prevent water from
traveling to the laser head and result in a nonfunctional laser.

Solution

Procedure

1. Tilt the power supply of the laser 30 degrees right or left.

2. Allow the power supply to remain in that position until the bubbles clear.

3. Try to start the laser again.

Spectrometer Troubleshooting

Continuous Beeping from System

Problem

The system is continuously beeping.

Probable Cause

The PC cannot communicate with the USB devices.

Solution

Procedure

1. Remove the USB devices from the PC,

2. Remove all installed Ocean Optics USB drivers, and

3. Restart the system.

4. Ensure that OOILIBS is installed, then reconnect the LIBS2000+ Spectrometer to the PC.