2: Installing the LIBS System
Note
If there are any yellow or red symbols next to each spectrometer entry, then you must
reinstall OOILIBS, reboot the system, and then reconnect the spectrometer.
3. In the OOILIBS program, select Tools | Select Spectrometer Modules. The HR2000
Configuration screen appears. Of the 8 possible spectrometer assignments, only 7 are active.
4. Click on the arrow for the first box (Unit A) to see a selection of 7 serial numbers.
5. Double-click on the lowest number and assign it to the first space. This is important in that it tells
the software which serial number correlates to the lowest wavelength spectrometer. The range
box should display the range of the spectrometer (this feature is not fully implemented).
Note
If you mistakenly assign an incorrect serial number, click on the text in the Unit Serial
Number column for that serial number. Highlight the text and delete it. Then, reassign the
correct serial number.
6. Proceed to assign serial numbers (in order of ascending serial numbers) to Units B-G.
7. Click the OK button when you have assigned all serial numbers.
8. You have now configured the Select Spectrometer Modules screen.
9. Exit OOILIBS software.
10. Restart OOILIBS software. The range at the bottom of the screen should be about 200–980 nm. If
this is not the case, or if you receive another error message, refer to Chapter 4:
Otherwise, you are ready to take data.