5: Calibration

Procedure

1. Create one VIS calibration and label the file with the spectrometer’s serial number and _VIS.

Then, create one UV calibration and label the VIS calibration file with the spectrometer’s serial
number and _UV. The extension .cal is added automatically to both files.

2. On

the

Irradiance Calibration screen, select Spectrometer | Combine UV/VIS Calibrations

from the menu bar. The Combine UV/VIS Calibrations screen appears.

3. On

the

Combine UV/VIS Calibrations screen, select File | Open UV Calibration from the

menu bar. The name of the UV calibration file that you just selected displays.

4. On

the

Combine UV/VIS Calibrations screen, select File | Open VIS Calibration from the

menu bar. The name of the VIS calibration file that you just selected displays.

5. All other default values should be OK, so select Spectrometer | Combine UV/VIS from the

menu bar.