5: Calibration

Use this screen to set the following:

a. For non-USB devices, enter the spectrometer number in the Spectrometer # field. (This

application reads the serial number from a USB2000 device.)

b. Adjust

the

Scans to Avg to some high number, depending on the integration period. For

example, if integration period is set to 2000ms, keep the number of scans to average at
around 30. This will make the response time around one minute per scan. If integration period
is much shorter, use a higher value to average.

10. Select Lamp | Select Lamp (file) from the menu bar and navigate to the appropriate lamp

calibration file to match the calibration lamp and the distance to be used. This file’s data is
displayed in the Calibration Lamp graph. Note that there are two files with the LS-1-CAL light
source, one for bare fiber and one for the CC-3 cosine corrector.