Timing tab – Ocean Optics OOIIrrad-C User Manual

Page 41

background image

4: Configuration

3. Enable

the

Log Calculated measurements every scan check box for automatic logging of all

active measurements using external triggers, if desired. Then enter the starting index number for
the log file.

4. Navigate to the appropriate file name and path to store full spectra.

5. Enable

the

Save spectrum file every scan check box for automatic saving of full spectra using

external triggers, if desired. Then enter the next serial number for the saved file.

6. Click

OK.

Timing Tab

The Timing tab on the Configure Spectrometer screen allows you to activate a preset duration for
experiments and a preset sampling interval.

Procedure

1. Enable

the

Preset Duration check box to set the amount of time for experiments. Then enter the

desired amount of time in the format Days | Hours | Minutes.

2. Enable

the

Preset Sampling Interval check box to set a sampling interval. Then enter the desired

amount of time for the interval in the format Hours | Minutes | Seconds.