Ocean Optics OOIIrrad-C User Manual
Page 24
3: Controls and Menus
Menu Shortcut
Keys
Description
Add Header to Log File
Adds an information header to the log (.txt) file defined in
the Configure Spectrometer window. Logging is done
with the Log Now button after log options are set. When
you click this button, a dialog box appears for optional
text to be logged with the data. The log file is a text-
based (ASCII) file that can be accessed by any
spreadsheet or text editing application.
Page Setup
Displays the Page setup dialog box for you to select
whether to print the header (name/date/page #), and
what page margins to use. You can select to use the
default margins, or enter custom margins.
Ctrl + P
Prints the currently displayed graph.
Exit
Exits the OOIIrrad-C application.
Lamp
Provides functions for calibrating the system’s lamp.
Current Calibration
Lamp’s Graph
Ctrl + Shift + L
Displays the graph for the current calibration lamp (must
be in uW/cm^2/nm mode) on the OOIColor/ OOIIrad
screen. This is useful for calibration verification. Overlay
the calibration lamp’s data from file over a scan of
calibration lamp just after calibrating the spectrometer
using the same lamp. Both should match.
Normalized
Displays the View actual lamp data or blackbody
curve
dialog box to enter the color temperature and the
value for normalization. The Lamp, normalized screen
then appears with the requested graph.
Enable Strobe
Ctrl + N
Enables or disables the spectrometer strobe control
function by toggling the spectrometer’s S0 line. A √ mark
appears by the menu item when the strobe is enabled.
Useful for turning compatible lamps on and off.