5: Calibration

Note

The Lamp | Enable Strobe function is used to control a digital line on the DB15
connector. It can be used to turn the lamp on and off.

11. Verify that the number of Scans to Avg is set to a high value such as 50 or 100 to ensure better

calibration accuracy.

12. Block the light path and select Spectrometer | Scan Dark to take a dark scan. The Store a scan

screen for a dark scan appears. Follow the directions on the screen. A raw baseline curve is added
to the Actual Scan graph.

13. Unblock the light path to calibration lamp and select Spectrometer | Scan Calibration Lamp to

take a reference scan. The Store a scan screen for a reference scan appears. Follow the directions
on the screen. The lamp’s raw spectrum, as seen by spectrometer, is added to the Actual Scan
graph. Click OK.