VEGA VEGAPULS 45 VBUS User Manual
Page 67
VEGAPULS 42, 44 and 45 – VBUS
67
Adjustment with the PC on the signal conditioning instrument
• Choose in the line "
Indication value" the
output "
Scaled" and e.g. the measured
value will be displayed (in liters and volume
percent), as well as the actual signal cur-
rent in the 0/4 … 20 mA signal cable.
Simulation
• Click to the menu "
Diagnostics/Simulation"
and choose the measurement loop (in the
example only one is available).
You are informed about the effect of the simu-
lation on relay or current outputs.
The menu window "
Simulation of outputs",
similar to the previous menu window, opens.
In this menu window however, you can set
the filling of the vessel or the signal current
and the indication to any value (simulate
measured value).
First of all, the actual measured value and the
signal current are displayed.
• Click to
"Start" in the turquoise window
segment.
The grey scroll bar becomes active. With this
scroll bar you can modify the measured
value in the range of -10 % … 110 % and
thereby simulate the filling or emptying of the
vessel. In the field of the turquoise window
cutout you can enter any % value for the
filling percentage.
Note:
The simulated measured value is outputted
during adjustment with the PC until you termi-
nate the simulation mode.