EXPLORER BALANCES

EN-91

5.2 Main Menu

The Explorer balance Main menu structure is illustrated below.

Calibration

User

Settings

Balance

Setup

Application

Modes

Weighing

Units

GLP and

GMP Data

Communication

Library

I/O

Settings

Diagnostics

Factory

Reset

Lockout

Information

5.3 Calibration

Explorer Balances offer a choice of six calibration methods: Internal Calibration, Automatic Calibration
(AutoCal), Calibration Adjust, Span Calibration, User Calibration and Calibration Test.

ExCal models have three calibration methods: Span Calibration, User Calibration and Calibration test.

Note: Do not disturb the balance during any calibration.

5.3.1

Calibration sub-menu

AutoCal™

Internal

Calibration

Automatic

Calibration

AutoCal™

Adjustment

Span

Calibration

User

Calibration

Calibration

Test


5.3.2

Internal Calibration (AutoCal

™)

Calibration is accomplished with the internal calibration mass. Internal calibration can be performed at
any time, provided the balance has warmed up to operating temperature and is level.

With the Balance turned ON and no load on the pan, touch AutoCal

Internal Calibration.

The Balance begins to calibrate.

Alternately, touch the shortcut Calibration Icon from any Application home screen
to initiate an Internal Calibration.

The display shows the status, then returns to the current application.

To cancel at any time, press Cancel.

5.3.3

Automatic Calibration

When Automatic Calibration is set ON, the balance performs a self calibration:

• when it senses a temperature change of 1.5°C

or every 11 hours

AutoCal will automatically calibrate the Balance (using the internal mass) each time there is a change in
temperature significant enough to affect accuracy.