EXPLORER BALANCES

Dynamic Weighing Application Setups (defaults in Bold)

Item

Available Settings

Comments

Enabled

On, Off

Turns Application on/off

Auto Sample Mode

Manual,
Semi Automatic
Automatic

To select the Mode

Statistics

Off, Manual, Auto

Lock Settings

On, Off

Press to lock the current setting(s)
so they cannot be changed.

Reference fields

Averaging Time

On, Off

Time in Seconds

Gross

On, Off

Gross weight is displayed

Net

On, Off

Net weight is displayed

Tare

On, Off

Tare weight is displayed

Print Options

See Printing Section.

Averaging Time

On, Off


4.5.3

Dynamic Weighing

– Semi-Auto


The cycle begins when an object (animal) is placed on the pan, but the held value must be reset manually.

To enable this feature, touch the Setup Icon (wrench) in the
top-center of the main screen to enter the
Dynamic Weighing Setup menu.

Change the mode to Semi Automatic.

The Dynamic

– Semi-Automatic screen appears with the

default (or last) Averaging Time displayed.

Place specimen
on the Pan. The
cycle starts
automatically.

After the cycle
ends, the weight
reading is held.
Press the

Reset

button to clear the
reading and return
to zero.