EXPLORER BALANCES

4.2

Parts Counting

Use this application to count samples of uniform weight. Three different modes can be selected:
Standard Counting, Check Counting, or Fill Counting.

4.2.1

Standard Parts Counting (default)

In the upper left portion of the home screen, select Counting.
The default (or last) Average Piece Weight (APW) is displayed.

Place objects on the pan to display the number of pieces.

The PARTS COUNTING-

Standard Home screen

Main Display Line

Second Display Line

Reference Fields

Application Buttons

Functions

Application Icon

Auto Optimization improves counting accuracy:
it re-calculates the piece weight automatically as
parts are added.

By default Auto Optimization is Off.

4.2.1.1 Edit Settings: To view or adjust the current settings, touch the Edit Settings button:

The Settings screen is now displayed.

Settings available: APW, Sample Size,

Functions available: Recalculate APW,
Return to Application

Establish an Average Piece Weight (APW)

Each time a new type of part is counted, the nominal weight of
one piece (Average Piece Weight or APW) must be established
using a small quantity of pieces. This APW is stored until
replaced by another APW.

There are two methods to establish the APW value:

1. The actual APW is known; or

2. The APW must be derived by weight. For this case the

current sample size will be used


To choose a different value, establish it beforehand, using the Sample Size button. (See below.)