M&C TechGroup CSS-VC2 Data sheet User Manual

Gas conditioning unit series css, Special features, Application

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

14.1a

Embracing Challenge

ƒ

ƒ

Compact high-performance construc-
tion with compressor cooler

ƒ

ƒ

Individually configurable, also beyond
the scope listed here

ƒ

ƒ

Completely pre-installed

ƒ

ƒ

Multi functional expandable control unit
TCU with warning and fault indicator in
plain text

ƒ

ƒ

One or two gas streams

ƒ

ƒ

19”- or wall mounting housing or port-
able in transport case

ƒ

ƒ

Low maintenance and self-monitoring

ƒ

ƒ

Outlet dew point adjustable from +2 °C
to +7 °C

ƒ

ƒ

Dew point stability ± 0,1°C

ƒ

ƒ

Operational in 10 minutes

ƒ

ƒ

Light weight

ƒ

ƒ

Optimum reliability

ƒ

ƒ

Operating hours counter

ƒ

ƒ

Resettable service interval

ƒ

ƒ

History memory

Special Features

Gas Conditioning Unit series CSS®

Version CSS-VC1 for max. 1 x 250Nl/h gas flow rate,
Version CSS-VC2 for max. 2 x 150Nl/h gas flow rate,
for 19”- or wall mounting or portable in transport case

CSS-VC1

Application

The units CSS-VC.. provide a complete pre-

installed compact high-performance sample

gas conditioning for continuous use. Due to
the multi functional expandable control unit
TCU and a big variety of additional options

the gas conditioning units excellently can be

adapted to the most diverse requirements of

gas analysis technique and therefore assume

tasks of complete anlayse systems.
Its compact construction only takes up little

space. The CSS-VC units are ready for use in

a few minutes. This makes time-consuming-

procurement of individual components and

assembly superfluous. Also a portable version

in transport case is possible.


Description

All components of the gas conditioning unit

are mounted in a robust compact sheet steel

case for wall- or 19»-rack mounting. The ven-

tilation grids in the sidewalls ensure that the

equipment is sufficiently ventilated.
Filter, flowmeter and peristaltic pump are

placed into the front panel and thus assure a

very easy maintenance. Due to a removable

lid and a foldable front plate also quick and

easy inspection and maintenance of all other

built-in components, especially easy change

of heat exchangers is possible.

The compressor gas cooler is equipped with

one or two heat exchangers of Duran glass,

stainless steel SS316Ti or PVDF.
Peristaltic pumps type SR25.2 are used for

contiuous condensate removal or also used

as dosing pump.
The current cooler temperature is displayed

at the front panel mounted multi functional

control unit TCU. Via keys the set temperature

of the cooler can be adjusted. In case of warn-

ings or faults in plain text are displayed and

additionally via LED’s indicated ( cooler within

set temperature range - green LED, flow alarm

(option) - yellow LED, temperature +-3°C from

set point and liquid alarm (option) - red LED.

The fine filter downstream cooler (different

filter executions possible) ensures the neces-

sary solid separation. The condition of the

filter element is visible from outside.
For protection of downstream analysers

against liquid breakthrough and for increase

of operating safety of the whole system

depending on the chosen type of filter a

suitable liquid alarm is possible or already

integrated in the filter.
A collective alarm message or on request

single alarm message inclusive mA outlet

for cooler temperature connected to output

terminals is available. Alarm message and

switching off the sample pump if present

happens via potential free contacts for collec-

tive alarm resp. single alarms (cooler tempera-

ture, flow, liquid in the system).

The downstream installed sample gas pump

N3/5/9 KPE is available with 3 pump capaci-

ties.
The flow meter FM40 with corresponding

measuring range installed in the sample gas

outlet can be equipped with a flow alarm

sensor FA-20mo and the adequate electronic.

Additionally for each sample outlet a second

sample outlet resp. bypass, also with flowm-

eter and flowalarm as option can be set up.
In case of aerosol problems a liquid particle

filter can be installed downstream the flowm-

eter in the sample gas outlet.
Via extension modules for control unit TCU

two more temperature controllers e.g. for

control of heated sample lines or gas sample

probes are available. Also external sample

gas pumps can be switched and gas sample

probes individually be purged back

.