M&C TechGroup SP2006-H280_DIL Data sheet User Manual

Gas sample probe series sp, Special features, Application

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

5.2

Embracing Challenge

ƒ

ƒ

Based on probe SP2000-H/....

ƒ

ƒ

Electrically heated up to 180 °C or 280 °C

ƒ

ƒ

NEMA 4, FRP housing

ƒ

ƒ

External heated critical orifice with addi-
tional safety filter

ƒ

ƒ

Including a dilution gas pre-heater

ƒ

ƒ

No dew point problems

ƒ

ƒ

Dilution ratios from 10:1 up to 500:1

ƒ

ƒ

With check gas connection

ƒ

ƒ

Independent of the process temperature

ƒ

ƒ

Changing of dilution and bypass block
without any tools within a few minutes

ƒ

ƒ

No disconnection of any tubes or hoses
for changing of dilution or bypass block

ƒ

ƒ

Easy operation and maintenance

Special Features

Gas Sample Probe Series SP®

Versions SP2006-H/DIL... and SP2006-H280/DIL
Gas Sample Dilution Probe

SP2006-H/DIL...

Application

The electrically heated M&C dilution probe

is applicable for such processes, where the

measuring procedure or the handling of the

process gas require the dilution of the sample

respectively the component to be measured.
Typical applications are the use with toxic

gases, the moisture measurement and the

emission measurement in flue gases.
The M&C dilution probe SP2006-H/DIL... is

based on the well tested M&C gas sample

probe SP2000-H. This enables the user to

adapt the probe to nearly every application,

using for example special filter techniques or

special materials.

Description

The dilution unit including the critical orifice

is mounted directly at the probe body and

therefore heated up with the probe to a sta-

ble temperature. The incorporated pre-heater

heats up the dilution gas to the probe tem-

perature. Both steps avoid that sample gas

decreases under the dewpoint. Calibration

gas enters the probe via an integrated con-

nection. The probe can be heated up to 180

°C or 280 °C. As an option a heated manual

operated 2 way ball valve is available. It is inte-

grated in the sample gas inlet and shuts off

the probe head from the process, for exam-

ple during the change of the filter element.

Additional the option ‘blow back in front of

ball valve’ via the probe flange is available.
Through an optional high flow rate check

valve /BB, which is fixed to the internal probe

area, blow back of the filter housing area and

the insitu probe tube will be done.Through

the optional high flow rate check valve /BB/F,

which is fixed to the filter housing wall, blow

back of the incorporated ceramic filter will be

done incl. the filter area and the insitu probe

tube. The optional isolation valve /I shuts off

the sample outlet from the internal filter area.

To optimize the blow back, an incorporated

accumulator tank with a volume of two lit-

ers is available as option. To drive this tank

an optional air actuated or solenoid valve is

offered.
The necessary dilution gas pressure can be

adjusted by the precision pressure controller

with pressure gauge.
A vacuum pressure gauge indicates the func-

tion of the dilution system.
The pressure controller and pressure gauges,

necessary for the dilution function, have to

be ordered separately. The external control

unit -S is equipped with an additional shut-

off ball valve and a flowmeter to adjust the

test gas quantity. The set -S1 includes an

additional pressure controller for the option

bypass injector -B or -BR.
The dilution probe realises dilution rates in

a range of 10:1 to 500:1. A high dilution rate

causes a low sample flow from the process.

With underpressure conditions the response

time can be minimised using the optional

heated bypass injector -B which is integrated

directly upstream the dilution unit. It is avail-

able with recirculation into the process -BR

as well.
The construction of the probe guarantees

an operation which is independent from the

process temperature and is easy to maintain

as well.