beautypg.com

Receipt of goods and storage, Illustration 1, Construction pump head – M&C TechGroup MP47-Z-BPR150 Operator's manual User Manual

Page 9

background image

9

Gas sampling and gas conditioning technology

6-1.1.7-ME

Voltage

230V 50Hz

115V 60Hz

230V 50Hz

115V 60Hz

Protection class

IP 20 - DIN 40050

Pump capacity max.

6,6 l/min* in the range of adjustment (outlet pressure)

Operating pressure
max.*

1,1 bar abs.

Operating pressure
min.**

0,46 bar abs. for 250Nl/h / 0,62 bar abs.

for 400Nl/h (outlet pressure 1,20 bar abs.)

Outlet pressure

1,15

– 1,3 bar abs. / at works 1,20 bar abs.

Gas temperature

-30 to +80°C

Ambient temperature

-10 to +40°C

Storage temperature

-15 to +60°C

Power consumption

110W

Current consumption

0,75A

1,7A

0,75A

1,7A

Gas connections

G1/4" i

DIN ISO 228/1

Electrical standard

EN 61010 Teil 1

Material of sample con-
tacting parts

PTFE, PVDF, Viton

PTFE, PVDF, Kalrez

Weight

4,6 Kg

Material mark according to ISO 1629 und 1043.1

* Maximum operating pressure for a constant outlet pressure and as a result a constant flow rate of the mea-

suring gas.

** The minimum operating pressure that still guarantees a constant pump capacity is changing according to the

necessary gas volume flow adjusted behind the pump and to the outlet pressure adjusted on the pressure
regulator. The lower the necessary gas volume flow and the outlet pressure, the lower is the minimal possi-
ble operating pressure that still guarantees a constant outlet pressure and therefore a constant sample gas
volume flow.

Illustration 1 Construction pump head


10

RECEIPT OF GOODS AND STORAGE

• Remove carefully the diaphragm pump and any accessories from the packaging and check im-

mediately against the delivery note whether the consignment is complete.