beautypg.com

M&C TechGroup PSS-5_3 Data sheet User Manual

Portable gas conditioning unit series pss, Special features, Application

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

14.5

Embracing Challenge

ƒ

ƒ

Low maintenance and self-monitoring

ƒ

ƒ

Outlet dew point adjustable from +2 °C
to +15 °C

ƒ

ƒ

Dew point stability ± 0,1 °C

ƒ

ƒ

Operational in 10 minutes

ƒ

ƒ

Compact construction

ƒ

ƒ

Light weight

ƒ

ƒ

Optimum reliability

ƒ

ƒ

Jet Stream heat exchangers in 3 stand-
ard materials

ƒ

ƒ

Universal equipment possible

Special Features

Portable Gas Conditioning Unit
series PSS®

Version PSS-5 for 150 Nl/h gas flow rate
Version PSS-5/3 für 350 Nl/h gas flow rate

PSS-5/3

Application

The M&C portable gas conditioning unit

PSS-5... is designed to carry out precise

gas analyses at different locations. The entire

gas conditioning unit is housed in a robust

aluminium-frame case to enable the user to

carry out a rapid gas analysis, which is safe

with little maintenance work.
The PSS-5... gas conditioning unit is suitable

for variable, discontinuous use as well as for

continuous operation. The components built

into the PSS-5... gas conditioning unit can be

used for standard applications. If you have

special problems please ask us for other solu-

tions.

Description

The PSS-5... gas conditioning unit is

equipped with an ECP... gas cooler

with Jet-Stream heat exchanger which

cools the sample gas to constant +5 °C

independent of the ambient temperature.

As soon as the operating temperature of

< +8 °C is reached after start-up, the gas

pump N...KPE is switched on automatically via

the status contact of the gas-cooler. The SR

25.1 peristaltic pump ensures a constant con-

densate removal which makes a long-term

measurement possible without problems.

The corresponding particle filtration is carried

out by a 2µm filter type FP-2T.
This makes the PSS5... unit a complete gas

conditioning system suitable for most of the

analysers.

Technical specifications and illustrations are without

obligation, subject to modifications. 09.95/12.11