2.16

Technical specifications and illustrations are without

obligation, subject to modifications. 11.09

∆P and T

90

at flow:

100

200

500

1000

Nl/hr

∆P Pressure loss SP30-H../EX2 length 1m

< 1

< 1

< 1

< 1

mbar

∆P pressure loss SP30-H1.1-V/EX2
length 1m with new pre-filter V20-2/30

< 1

< 1

1,5

4

mbar

T

90

- time

14

7

< 3

< 2

sec

Pressure difference and T90-time

Temperature controller type

HEX5-0.2.08

Ex-marking

II 3 G Ex nA nC nL IIC T5

Ambient temperature

0°C to +50°C

Power supply

230V 50/60Hz max. 800W or 115V 50/60Hz max. 830W

Electrical connection

Cable gland, terminal range 6-12mm, terminals max. 4mm

2

Electrical standard

EN 61010, EN60519-1

Degree of protection

IP65 EN 60529

Temperature status alarm

-10°C to T

set

switching capacity 250V 3A AC, 0,25A DC potential free

Excess temperature limiter

+5°C to T

set

manual reset

Mounting

Wall mounting

Temperature class

Operating/sampling temperature °C

Limiter °C

T2 (only with thermocouple Fe-CuNi)

280

290

T3

185

190

T4

120

125

T5

85

90

Temperature classes