[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.16
Technical specifications and illustrations are without
obligation, subject to modifications. 11.09
Heated sample tube Series SP
®
Type
SP30-H1.1/EX2
SP30-H2/EX2
SP30-H1.1-V/EX2
Temperature sensor
Fe-CuNi
PT100 2 wire
Fe-CuNi
Temperature controller
external
Probe tube length L1
max. 1 m
Sample temperature max.
depending on the temperature class (see table temperature classes). Please specify with order.
Operating temperature max.
depending on the temperature class (see table temperature classes). Please specify with order.
Pre-filter
optional
optional
V20-2/30 insiti filter
length 520 mm, ø 60 mm,
filter porosity 2µm, integrated
and heated
Sample gas inlet connection
G3/4"i DIN ISO 228/1
pre-filter with G1 1/2"i DIN
ISO 228/1
Dust loading
max. 2 g/m
3
>2 g/m
3
Probe tube volume
380 ml/m
420 ml
Sample pressure max.
5 bar g
Storage temperature
-30 °C to +70 °C
Ready for operation
2 h
Power supply
230V 50/60Hz or 115V 50/60Hz
Heating capacity
0,6m: 600W, 1m: 800W (other length/capacities on request)
Electrical connections
terminals, max. 4 mm
2
, 2x PG13,5 cable gland, terminal range 6-12mm
Electrical standard
EN 61010, EN60519-1
Degree of protection
IP54 EN 60529
Ex-marking
II 3 G Ex nA II T5 to T2 see table temperature class
Ambient temperature
connection box of the heated sample tube: -20 to +70°C
Mounting flange
DN65 PN6, Form B with 4 welded screws on both sides M 12x 40 mm
Material of parts in contact with the sample
stainless steel 1.4539
1.4539, SS316Ti/SS316
Technical Data
210
120
Ø 130
20
∅ 160
14 40
M12
L1
Terminals
Ø 42,2
G3/4"
PG13,5
Temperature-Sensor
Power
Temperature of pre filter in dependence of controller
temperature at 20°C ambient temperature.
500
ш 64
210
120
ш 130
ш 160
14 40
M12
L1
Terminals
PG13,5
Temperature
Sensor
Power
Controller temperature
180°C
Temperature of pre filter
135°C
Probe tube SP30-H1.1/EX2 and SP30-H2/EX2
Dimensions in mm
Probe tube with pre-filter SP30-H1.1-V/EX2
Dimensions
9
1
12 11
6
4
2 3
5
14
7 8
10 11
13
12
14
16
15
A1 A2
SICHERUNG NICHT
UNTER SPANNUNG
HERAUSNEHMEN!
DO NOT REMOVE
FUSE WHEN
ENERGIZED!
Sicherung 1A
Fuse 1A
175
250
145
200
225
120
Ø
6,5
Temperatue controller HEX5-0.2.08
Dimensions in mm